m = p1 / ( 1 + (nu/p2)^2 )^p3 + (1-p1) / ( 1 + (nu/p4)^2 )^p5
Legend: MTF measurements of Loral-Lesser CCD @670nm. Plusses are the measurements, diamonds the 'silicon MTF' after dividing with the pixel MTF (SINC(nu) = SIN( pi * nu ) / ( pi * nu ), shown as a dotted curve). The dashed curve gives the 'Moffat' fit to the silicon MTF.
With the fit in hand, we can invert the MTF into a line spread function (LSF) through a Fourier transform. The result is shown in the figure below. The sharp peak in the silicon LSF is emerging as a result of extrapolating the silicon MTF to high frequencies. This does not change the real detector LSF, but effectively only affects the slope of the small 'jump' at the pixel edge. Having the one dimensional silicon MTF, the 2D silicon and detector point spread functions can be derived, if it is assumed that the silicon MTF is rotationally symmetric.
| 543nm | 670nm | 830nm | |
|---|---|---|---|
| Pixel Strehl ratio | 0.199 | 0.291 | 0.434 |
| Effective pixel size | 2.24 pix | 1.85 pix | 1.52 pix |
| 80% Ensquared Energy | 3.90 pix | 3.51 pix | 3.33 pix |
| 80% Encircled Energy | 4.21 pix | 3.74 pix | 3.48 pix |
The silicon-PSF at the 3 wavelengths are remarkably different, as can be seen from the left-hand figure above, with the 543nm silicon-PSF having no sharp core at all. The encircled/ensquared energy is plotted in right-hand figure above and listed in the following table. It is obvious that the extended wings of the PSF's are important at all wavelengths. Note that the EEs cross eachother at a diameter of 6-7 pixels. A corresponding trend in the raw MTF data is present, where there is an apparent cross over at ~0.15cy/pixel, the 543nm MTF being the better at the lowest frequencies.
| Encircled Energy | Ensquared Energy | |||||
|---|---|---|---|---|---|---|
| Box size | 543nm | 670nm | 830nm | 543nm | 670nm | 830nm |
| [pixels] | [%] | [%] | [%] | [%] | [%] | [%] |
| 1 | 16.08 | 25.75 | 41.48 | 16.51 | 24.27 | 38.80 |
| 2 | 43.27 | 50.40 | 59.93 | 45.93 | 53.01 | 61.06 |
| 3 | 64.41 | 70.31 | 74.82 | 67.85 | 73.26 | 76.36 |
| 4 | 77.88 | 82.45 | 84.31 | 81.03 | 84.71 | 85.41 |
| 5 | 86.10 | 88.80 | 89.68 | 88.69 | 90.48 | 90.23 |
| 6 | 91.20 | 92.20 | 92.63 | 93.25 | 93.57 | 92.81 |
| 7 | 94.49 | 94.32 | 94.33 | 96.05 | 95.47 | 94.30 |
| 8 | 96.63 | 95.79 | 95.44 | 97.77 | 96.78 | 95.31 |
| 9 | 98.01 | 96.88 | 96.24 | 98.82 | 97.72 | 96.07 |
| 10 | 98.87 | 97.70 | 96.91 | 99.42 | 98.41 | 96.73 |
| 11 | 99.39 | 98.32 | 97.49 | 99.75 | 98.92 | 97.29 |
| 12 | 99.69 | 98.79 | 98.02 | 99.92 | 99.29 | 97.85 |
Approximately 8% of the flux seems to be absorbed within the well at 670nm while it is ~25% at 870nm. Using an absorption coefficient of 0.067/mu at 830nm and 0.18/mu at 670nm, this indicate a field free region of ~15mu.
The graph above shows the effect on MTF from changing the P1, P2
voltages. The MTF is measured at one spatial frequency only (~0.4 cycle/pixel),
and is
normalized to the MTF in MPP mode, represented with the single
data point outside the linear curve.
The gain in MTF appears to be proportional to the voltage applied to P1, P2.
Using the highest P1, P2 voltage of +12V, which gave the best MTF in the previous graph, the voltage of P3 is now changed. The normalization is the same as before. The effect on MTF is slight, but with P3 above -3V, a linear increase with voltage is seen.
As a relative improvement of 43% in MTF was reached by changing the
phase voltages from the standard values was reached, the method
seems very attractive.
There are some drawbacks, though:
The higher phase voltages means more dark current. At the highest voltages
used, the dark current was about 30e-/h, approximately a 20-fold increase
compared to the MPP mode. Still, for several applications, this is an
acceptable level.
In the standard setup, the full well is 115.000e- for the CCD tested,
regardless of MPP mode. By increasing the P1, P2 voltages, no harm
to the well capacity was observed, but increasing P3 also caused
problems: At voltages P1=+12V P2=+12V P3=+6V, the full well was only
16.000e-. At higher P3 voltages, the imaging capability was lost.
At a P3 voltage of +2V, the full well was 69.000e- and the relative
gain in MTF was 38%, which seems like a useful compromise.
The high voltages would cause problems with spurious charge and CTE
during pre-clear and read-out. To avoid this, standard voltages were
used at these times.
The typical appearance of "cosmic hits" in a dark exposure, using a
thinned Loral 2k CCD from the Copenhagen lot.
To the left a hit in a shallow angle appears comet-like, as charge
diffusion is large at the back side and no diffusion occurs near the
electrodes.
To the right, the stellar-like image from a hit approximately
perpendicular to the CCD plane.
The spreading of charge from a particle impact is not only due to diffusion.
The large number of charges in a small area also causes self-repulsion.
Last updated September 28, 1999